Numerical verification of the hydrodynamic reconstruction method for contact resonance atomic force microscopy

Loading...
Thumbnail Image

Authors

Shihab, Rafiul
Tung, Ryan C.

Issue Date

2018

Type

Article

Language

Keywords

Research Projects

Organizational Units

Journal Issue

Alternative Title

Abstract

A numerical verification of an experimental method used to estimate hydrodynamic forces in contact resonance atomic force microscopy (CR AFM) is performed. The experimental estimation technique, known as the Hydrodynamic Reconstruction Method (HRM), is verified for three distinct cantilever geometries at several vibrational eigenmodes and sample stiffnesses. The results of the analysis are discussed and recommendations for the applicable measurement range of the HRM are provided. (C) 2018 Author(s).

Description

Citation

Shihab, R., & Tung, R. C. (2018). Numerical verification of the hydrodynamic reconstruction method for contact resonance atomic force microscopy. AIP Advances, 8(8), 085015. doi:10.1063/1.5044651

Publisher

License

Creative Commons Attribution 4.0 International